Ex vivo assessment of erythrocyte tolerance to the HeartWare ventricular assist device operated in three discrete configurations
Kuck, L., Simmonds, M. J., Chan, C., Pauls, J. P., Tansley, G. D., Feldmann, F., & McNamee, A. P. (2021). Ex vivo assessment of erythrocyte tolerance to the HeartWare ventricular assist device operated in three discrete configurations. Artificial organs, 45(6), E146–E157. https://doi.org/10.1111/aor.13877
Abstract: Despite technological advances in ventricular assist devices (VADs) to treat end-stage heart failure, hemocompatibility remains a constant concern, with supraphysiological shear stresses an unavoidable reality with clinical use. Given that impeller rotational speed is related to the instantaneous shear within the pump housing, it is plausible that the modulation of pump speed may regulate peak mechanical shear stresses and thus ameliorate blood damage. The present study investigated the hemocompatibility of the HeartWare HVAD in three configurations typical of clinical applications: standard systemic support left VAD (LVAD), pediatric support LVAD, and pulmonary support right VAD (RVAD) conditions. Two ex vivo mock circulation blood loops were constructed using explanted HVADs, in which pump speed and external loop resistance were manipulated to reflect the flow rates and differential pressures reported in configurations for standard adult LVAD (at 3150 rev⸱min-1 ), pediatric LVAD (at 2400 rev⸱min-1 ), and adult RVAD (at 1900 rev⸱min-1 ). Using bovine blood, the mock circulation blood loops were tested at 37°C over a period of 6 hours (consistent with ASTM F1841-97) and compared with static control.